Published 1974
| Version v1
Report
Open
On the operation of a cryostat for Ge(Li) detector
Creators
Description
Operating experience with cryostats for Ge(Li) detectors developed at the FTI of the Academy of Science of the Ukrainian SSR, and used in several laboratories for 5 years is reported. It is shown that the spectrometric properties of all cryostat-mounted detectors operating since 1969 have not been affected and up to now the detectors are operating successfully. Nitrogen consumption has not increased and is at a level of 0.5 l/d. During five-year continuous operation the cryostat pressure has varied from 6.10-7 to 8.10-6 torr
Files
8287612.pdf
Files
(9.3 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:6ed84456067b9bf043ba726c2ff00f20
|
9.3 MB | Preview Download |
System files
(290.0 kB)
| Name | Size | Download all |
|---|
Additional details
Additional titles
- Original title (Russian)
- К вопросу об эксплуатации криостата для Ге(Ли)-детектора
Publishing Information
- Imprint Title
- Nuclear science and engineering problems
- Imprint Pagination
- no. 1(2) p. 50-51.
- Series
- Fizika i tekhnika vysokogo vakuuma.
- Report number
- KFTI--74-24
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 8287612
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S75: CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND SUPERFLUIDITY;
- Descriptors DEI
- CRYOSTATS; HIGH VACUUM; LI-DRIFTED GE DETECTORS; PERFORMANCE; TIME DEPENDENCE
- Descriptors DEC
- CONTROL EQUIPMENT; GE SEMICONDUCTOR DETECTORS; LI-DRIFTED DETECTORS; MEASURING INSTRUMENTS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; THERMOSTATS
Optional Information
- Notes
- 2 refs.; 1 fig. Imprint:Voprosy atomnoj nauki i tekhniki.