On the operation of a cryostat for Ge(Li) detector

Description

Operating experience with cryostats for Ge(Li) detectors developed at the FTI of the Academy of Science of the Ukrainian SSR, and used in several laboratories for 5 years is reported. It is shown that the spectrometric properties of all cryostat-mounted detectors operating since 1969 have not been affected and up to now the detectors are operating successfully. Nitrogen consumption has not increased and is at a level of 0.5 l/d. During five-year continuous operation the cryostat pressure has varied from 6.10-7 to 8.10-6 torr