Patient dose in interventional radiology: a multicentre study of the most frequent procedures in France
Creators
- 1. French Society of Medical Physicists (SFPM), Paris (France)
- 2. Institut de Radioprotection et de Surete Nucleaire, Fontenay-aux-Roses (France)
- 3. La Timone University Hospital, Department of Radiology, Marseille Cedex (France)
- 4. Lariboisiere Hospital, Department of Medical Physics and Radiation Protection, Paris (France)
- 5. French Society of Radiology (SFR) - Interventional Radiology Federation (FRI), Paris (France)
- 6. Nimes University Hospital, Medical Imaging Group Nimes, Department of Radiology, Nimes (France)
- 7. Trousseau University Hospital, Department of Pediatric Radiology, Paris (France)
Description
A national retrospective survey on patient doses was performed by the French Society of Medical physicists to assess reference levels (RLs) in interventional radiology as required by the European Directive 2013/59/Euratom. Fifteen interventional procedures in neuroradiology, vascular radiology and osteoarticular procedures were analysed. Kerma area product (KAP), fluoroscopy time (FT), reference air kerma and number of images were recorded for 10 to 30 patients per procedure. RLs were calculated as the 3rd quartiles of the distributions. Results on 4600 procedures from 36 departments confirmed the large variability in patient dose for the same procedure. RLs were proposed for the four dosimetric estimators and the 15 procedures. RLs in terms of KAP and FT were 90 Gm.cm2 and 11 mins for cerebral angiography, 35 Gy.cm2 and 16 mins for biliary drainage, 75 Gy.cm2 and 6 mins for lower limbs arteriography and 70 Gy.cm2 and 11 mins for vertebroplasty. For these four procedures, RLs were defined according to the complexity of the procedure. For all the procedures, the results were lower than most of those already published. This study reports RLs in interventional radiology based on a national survey. Continual evolution of practices and technologies requires regular updates of RLs. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00330-017-4780-5Additional details
Identifiers
Publishing Information
- Journal Title
- European Radiology
- Journal Volume
- 27
- Journal Issue
- 10
- Journal Page Range
- p. 4281-4290
- ISSN
- 0938-7994
- CODEN
- EURAE3
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 48091656
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BENCHMARKS; BIOMEDICAL RADIOGRAPHY; BLOOD VESSELS; BRAIN; COMPUTERIZED TOMOGRAPHY; DOCUMENT TYPES; FLUOROSCOPY; FRANCE; KERMA; PATIENTS; PERSONNEL DOSIMETRY; RADIATION DOSES; RADIATION PROTECTION; SURGERY; X-RAY RADIOGRAPHY
- Descriptors DEC
- BIOMEDICAL RADIOGRAPHY; BODY; CARDIOVASCULAR SYSTEM; CENTRAL NERVOUS SYSTEM; DEVELOPED COUNTRIES; DIAGNOSTIC TECHNIQUES; DOSES; DOSIMETRY; EUROPE; INDUSTRIAL RADIOGRAPHY; MATERIALS TESTING; MEDICINE; NERVOUS SYSTEM; NONDESTRUCTIVE TESTING; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; TESTING; TOMOGRAPHY; WESTERN EUROPE