Improvements in radiographic imaging obtained with the Super Rotalix Ceramic X-ray tube

  • 1. Mueller (C.H.F.) G.m.b.H., Hamburg (Germany, F.R.)

Description

The Super Rotalix Ceramic X-ray tube is described, with particular reference to those characteristics that improve image quality mainly by reducing extrafocal radiation. A technique for measuring extrafocal radiation is described, and the results obtained with the Super Rotalix Ceramic are compared with those of a conventional X-ray tube. The reduction in extrafocal radiation is considerable, and the resulting improvement in image quality is greater than anticipated. It is concluded that incorporation of the Super Rotalix Ceramic in the X-ray imaging chain can yield a significant improvement in overall image quality. (Auth.)

Additional details

Publishing Information

Journal Title
Medicamundi
Journal Volume
25
Journal Issue
3
Series
Medicamundi.
Journal Page Range
118-121
ISSN
0025-7664

INIS

Country of Publication
Netherlands
Country of Input or Organization
Netherlands
INIS RN
12619108
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; IMAGE PROCESSING; OPTIMIZATION; X-RAY TUBES
Descriptors DEC
DIAGNOSTIC TECHNIQUES; ELECTRON TUBES; MEDICINE; X-RAY EQUIPMENT