MOSFET dosimetry: temperature effects in-vivo
Creators
- 1. City University of Hong Kong, Kowloon Tong (China). Department of Physics and Materials Science
- 2. Cancer Services, Wollongong, NSW (Australia). Department of Medical Physics
Description
Full text: This note investigates temperature effects on dosimetry using a Metal Oxide Semiconductor Field Effect Transistor (MOSFET) for radiotherapy x-ray treatment. This was performed by analysing the dose response and threshold voltage outputs for MOSFET dosimeters as a function of ambient temperature. Results have shown the clinical semiconductor dosimetry system (CSDS) MOSFET provides stable dose measurements with temperatures varying from 15 deg C up to 40 deg C. Thus standard irradiations performed at room temperature can be directly compared to in-vivo dose assessments performed at near body temperature without a temperature correction function. The MOSFET dosimeter threshold voltage varies with temperature and this level is dependant on the dose history of the MOSFET dosimeter. However the variation can be accounted for in the measurement method. For accurate dosimetry the detector should be placed for approximately 60 seconds on a patient to allow thermal equilibrium before measurements are taken with the final reading performed whilst still attached to the patient or conversely left for approximately 120 seconds after removal from the patient if initial readout was measured at room temperature to allow temperature equilibrium to be established. Copyright (2004) Australasian College of Physical Scientists and Engineers in Medicine
Additional details
Publishing Information
- Journal Title
- Australasian Physical and Engineering Sciences in Medicine
- Journal Volume
- 27
- Journal Issue
- 4
- Journal Page Range
- p. 328-329
- ISSN
- 0158-9938
- CODEN
- AUPMDI
Conference
- Title
- EPSM 2004. Regional healthcare technologists overcoming the tyrany of distance
- Dates
- 14-18 Nov 2004
- Place
- Geelong, VIC (Australia)
INIS
- Country of Publication
- Australia
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 37103413
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- AMBIENT TEMPERATURE; BODY TEMPERATURE; DOSEMETERS; DOSIMETRY; IN VIVO; MOSFET; PATIENTS; RADIATION DOSES; RADIOTHERAPY; SEMICONDUCTOR MATERIALS; TEMPERATURE DEPENDENCE; THERMAL EQUILIBRIUM
- Descriptors DEC
- DOSES; EQUILIBRIUM; FIELD EFFECT TRANSISTORS; MATERIALS; MEASURING INSTRUMENTS; MEDICINE; MOS TRANSISTORS; NUCLEAR MEDICINE; RADIOLOGY; SEMICONDUCTOR DEVICES; THERAPY; TRANSISTORS