The potential failure risk of the cone-beam computed tomography-based planning target volume margin definition for prostate image-guided radiotherapy based on a prospective single-institutional hybrid analysis

  • 1. Department of Radiation Oncology, Southern Tohoku BNCT Research Center, 7-10, Yatsuyamada, Koriyama, Fukushima, 963-8052 (Japan)
  • 2. Department of Radiology and Radiation Oncology, Hirosaki University Graduate School of Medicine, 5 Zaifu-cho, Hirosaki, Aomori, 036-8562 (Japan)
  • 3. Division of Radiology, Hirosaki University Hospital, 53 Hon-cho, Hirosaki, Aomori, 036-8563 (Japan)

Description

The purpose of this study was to evaluate the impact of markerless on-board kilovoltage (kV) cone-beam computed tomography (CBCT)-based positioning uncertainty on determination of the planning target volume (PTV) margin by comparison with kV on-board imaging (OBI) with gold fiducial markers (FMs), and to validate a methodology for the evaluation of PTV margins for markerless kV-CBCT in prostate image-guided radiotherapy (IGRT). A total of 1177 pre- and 1177 post-treatment kV-OBI and 1177 pre- and 206 post-treatment kV-CBCT images were analyzed in 25 patients who received prostate IGRT with daily localization by implanted FMs. Intrafractional motion of the prostate was evaluated between each pre- and post-treatment image with these two different techniques. The differences in prostate deviations and intrafractional motions between matching by FM in kV-OBI (OBI-FM) and matching by soft tissues in kV-CBCT (CBCT-ST) were compared by Bland-Altman limits of agreement. Compensated PTV margins were determined and compensated by references. Mean differences between OBI-FM and CBCT-ST in the anterior to posterior (AP), superior to inferior (SI), and left to right (LR) directions were − 0.43 ± 1.45, − 0.09 ± 1.65, and − 0.12 ± 0.80 mm, respectively, with R2 = 0.85, 0.88, and 0.83, respectively. Intrafractional motions obtained from CBCT-ST were 0.00 ± 1.46, 0.02 ± 1.49, and 0.15 ± 0.64 mm, respectively, which were smaller than the results from OBI-FM, with 0.43 ± 1.90, 0.12 ± 1.98, and 0.26 ± 0.80 mm, respectively, with R2 = 0.42, 0.33, and 0.16, respectively. Bland-Altman analysis showed a significant proportional bias. PTV margins of 1.5 mm, 1.4 mm, and 0.9 mm for CBCT-ST were calculated from the values of CBCT-ST, which were also smaller than the values of 3.15 mm, 3.66 mm, and 1.60 mm from OBI-FM. The practical PTV margin for CBCT-ST was compensated with the values from OBI-FM as 4.1 mm, 4.8 mm, and 2.2 mm. PTV margins calculated from CBCT-ST might be underestimated compared to the true PTV margins. To determine a reliable CBCT-ST-based PTV margin, at least the systemic error Σ and the random error σ for on-line matching errors need to be investigated by supportive preliminary FM evaluation at least once.

Availability note (English)

Available from http://dx.doi.org/10.1186/s13014-018-1043-9; Available from http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5992771

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Oncology (Online)
Journal Volume
13
Journal Page Range
vp.
ISSN
1748-717X

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
49082605
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COMPUTERIZED TOMOGRAPHY; IMAGE PROCESSING; PROSTATE; RADIATION HAZARDS; RADIOTHERAPY
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; GLANDS; HAZARDS; HEALTH HAZARDS; MALE GENITALS; MEDICINE; NUCLEAR MEDICINE; ORGANS; PROCESSING; RADIOLOGY; THERAPY; TOMOGRAPHY

Optional Information

Copyright
Copyright (c) The Author(s). 2018
Notes
PMCID: PMC5992771; PMID: 29880006; PUBLISHER-ID: 1043; OAI: oai:pubmedcentral.nih.gov:5992771