Published February 1983 | Version v1
Journal article

A simple equivalent tissue-air ratio method for calculating absorbed dose in a heterogeneous medium

  • 1. Rambam Medical Center, Haifa, Israel

Description

Density correction factors for 60Co gamma rays and 8-MV x rays were measured with an ionization chamber in water-equivalent phantoms containing simulated lung or bone inhomogeneities larger or smaller than the beam (beam sizes, 5 x 5 to 30 x 30 cm). Measurements were compared with values calculated from the effective SSD method, the generalized Batho power law method, and a simplified equivalent tissue-air-ratio (SETAR) method. The results indicate that the SETAR method seems adequate for most clinical applications

Additional details

Publishing Information

Journal Title
Radiology
Journal Volume
146
Journal Issue
2
Series
Radiology.
Journal Page Range
527-529
ISSN
0033-8419