Published November 2016 | Version v1
Journal article

Measurement of turbulent electron temperature fluctuations on the ASDEX Upgrade tokamak using correlated electron cyclotron emission

  • 1. Plasma Science and Fusion Center, Massachusetts Institute of Technology, Cambridge, Massachusetts 02139 (United States)
  • 2. Max Planck Institute for Plasma Physics, 85748 Garching (Germany)
  • 3. FOM Institute DIFFER, 5612 AJ Eindhoven (Netherlands)

Description

Turbulent temperature fluctuations are measured on the ASDEX Upgrade tokamak using pairs of closely spaced, narrow-band heterodyne radiometer channels and a standard correlation technique. The pre-detection spacing and bandwidth of the radiometer channel pairs is chosen such that they are physically separated less than a turbulent correlation length, but do not overlap. The radiometer has 4 fixed filter frequency channels and two tunable filter channels for added flexibility in the measurement position. Relative temperature fluctuation amplitudes are observed in a helium plasma to be δT/T = (0.76 ± 0.02)%, (0.67 ± 0.02)%, and (0.59 ± 0.03)% at normalised toroidal flux radius of ρtor = 0.82, 0.75, and 0.68, respectively.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
87
Journal Issue
11
Journal Page Range
vp.
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2016 Author(s)