Investigation of modulation transfer function (MTF) in a digital imaging system
- 1. Osaka City Univ. (Japan). Faculty of Medicine
Description
The modulation transfer function (MTF) has been used for evaluation of radiographic image quality. However, in digital radiographic system such as a computed radiography (CR), the MTF can not be used in the same way of an analog system of evaluation because digital system is not shift-invaliant system. Then, we suggest the 'MTFfinal' which includes geometric unsharpness, detector unsharpness, unsharpness of the sampling aperture and unsharpness of the display (Laser-printer). It is an analog component in the digital system and in agreement with the definition of the MTF. Furthermore, an overall aliasing has to be used with MTFfinal for evaluating the digital system because it is one of the important factors of the digital system. The MTFfinal and the overall aliasing can be obtained from the overall MTFs of the center alignment and the shifted alignment. (author)
Additional details
Additional titles
- Subtitle (English)
- Final MTF and overall aliasing
Publishing Information
- Journal Title
- Nippon Hoshasen Gijutsu Gakkai Zasshi
- Journal Volume
- 48
- Journal Issue
- 11
- Journal Page Range
- p. 1932-1938.
- ISSN
- 0369-4305
- CODEN
- NIPHAP
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 24026693
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; COMPUTERS; DIGITAL SYSTEMS; FREQUENCY MODULATION; IMAGES; TRANSFER FUNCTIONS
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; FUNCTIONS; MEDICINE; MODULATION