Dose metrology: TLD/OSL dose accuracy and energy response performance
Creators
- 1. University Abdelmalek Essaadi, Tetouan (Morocco)
- 2. Ibn Tofail University, Kenitra (Morocco)
- 3. ENA, Marrakech (Morocco)
- 4. National Center of Radiation Protection (CNRP), Sale (Morocco)
- 5. National Center for Energy, Science and Nuclear Technology (CNETSEN), Rabat (Morocco)
Description
An essential step in evaluating and comparing the performance of two passive radiation dosimeter types, thermoluminescent (TLD) and optically stimulated luminescence (OSL), used by workers in environments with ionizing radiation for individual radiological monitoring and control of external exposure at various times (cumulative dose for 1 month), is to compare the measured dose accuracy, energy response, and coefficient of variation. In fact this performance study consists in determining the accuracy of both R(10) and R(0.07) which are considered as the ratios of the measured dose (Hp(10) or Hp(0.07)) to the delivered dose (Hp(10) or Hp(0.07)) for each photon energy. The validity of the results of this test is based on the acceptance limits of the ICRP and the international standard IEC-62387. The relative energy response used is normalized to the 137Cs 662 keV energy to find which energy response is closest to the ideal case, and the coefficient of variation that allows to determine the statistical fluctuation of the Hp(10) and Hp(0.07) doses. The results of the accuracy test for the OSL and TLD dosimeters are acceptable because they fall within the ICRP limits. For the energy response, the OSL performs better than the TLD for Hp(10) and Hp(0.07), and for the coefficient of variation, the OSL satisfies the requirements of ISO 62387 for both Hp(10) and Hp(0.07), while the TLD satisfies these requirements only for the measurement of Hp (0.07)
Additional details
Publishing Information
- Journal Title
- Nuclear Engineering and Technology
- Journal Volume
- 55
- Journal Issue
- 2
- Journal Page Range
- p. 717-724
- ISSN
- 1738-5733
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 55095489
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ACCURACY; CESIUM 137; DOSES; ENERGY; FLUCTUATIONS; ICRP; IONIZING RADIATIONS; KEV RANGE; METROLOGY; MONITORING; PERFORMANCE; PHOTONS; STANDARDS; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BOSONS; CESIUM ISOTOPES; DOSEMETERS; ELEMENTARY PARTICLES; ENERGY RANGE; INTERMEDIATE MASS NUCLEI; INTERNATIONAL ORGANIZATIONS; ISOTOPES; LUMINESCENT DOSEMETERS; MASSLESS PARTICLES; MEASURING INSTRUMENTS; NUCLEI; ODD-EVEN NUCLEI; RADIATIONS; RADIOISOTOPES; VARIATIONS; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 21 refs, 14 figs, 4 tabs