Accurate determination of Ge(Li) spectrometer efficiency and practical correction for dead time and pulse pile-up
Creators
- 1. Ustav pro Vyzkum, Vyrobu a Vyuziti Radioisotopu, Prague (Czechoslovakia)
Description
The accurate adjustment is described of the semiconductor Ge(Li) spectrometer efficiency by means of 137Cs and 141Ce activity standards of the EFF type produced by UVVVR Prague. A comparison is made of three methods for determining analyzer dead time correction and the correction for count rate losses due to pulse pile-up. The spectrum acquisition proceeded in real time mode while the analyzer dead time and the pile-up effect corrections were made using a CI 104 R pulse generator. The 137Cs peak area determination was performed as a summation over the (-3 FWHM, +2 FWHM) region from which the linear background was subtracted obtained by fitting to adjacent regions of 1 FWHM in length. In testing the methods, the corrected 137Cs peak area was studied in dependence on the total count rate. This was altered within 600 and 50,000 s-1 by placing a 141Ce source to different distances from the detector. The 137Cs source was in a constant position. (B.S.)
Additional details
Additional titles
- Original title (Czech)
- Presne stanoveni ucinnosti Ge(Li) spektrometru a prakticke provedeni korekce mrtve doby a vlivu nakupeni impulsu
Publishing Information
- Journal Title
- Radioisotopy
- Journal Volume
- 18
- Journal Issue
- 4
- Series
- Radioisotopy.
- Journal Page Range
- 511-518
Conference
- Title
- Conference on instrumental activation analysis IAA '77.
- Dates
- 20 - 24 Jun 1977.
- Place
- Novy Bor, Czechoslovakia.
INIS
- Country of Publication
- Serbia and Montenegro
- Country of Input or Organization
- Serbia and Montenegro
- INIS RN
- 10452856
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; CALIBRATION STANDARDS; CERIUM 141; CESIUM 137; CORRECTIONS; DEAD TIME; EFFICIENCY; LI-DRIFTED GE DETECTORS; PEAKS; PULSE PILEUP
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CERIUM ISOTOPES; CESIUM ISOTOPES; DAYS LIVING RADIOISOTOPES; EVEN-ODD NUCLEI; GE SEMICONDUCTOR DETECTORS; INTERMEDIATE MASS NUCLEI; ISOTOPES; LI-DRIFTED DETECTORS; MEASURING INSTRUMENTS; NUCLEI; ODD-EVEN NUCLEI; RADIATION DETECTORS; RADIOISOTOPES; RARE EARTH NUCLEI; SEMICONDUCTOR DETECTORS; TIMING PROPERTIES; YEARS LIVING RADIOISOTOPES