Published August 11, 2003
| Version v1
Journal article
A gas analyzer for the internal polarized target of the HERMES experiment
Description
A gas analyzer has been developed for the internal polarized target of the HERMES experiment at DESY in order to determine the relative amount of atomic and molecular hydrogen or deuterium in a gas sample. The precise quantitative knowledge of this ratio is crucial because the nucleons in atoms and molecules contribute differently to the average nuclear polarization of the target gas. A new calibration technique used to derive the relative sensitivity to atoms and molecules is presented. As an example, it is shown how the gas analyzer is used within the HERMES environment to divide the molecules in the gas sample into an unpolarized and a potentially polarized fraction
Additional details
Identifiers
- PII
- S0168900203017029;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 508
- Journal Issue
- 3
- Journal Page Range
- p. 268-275
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 35021571
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; DESY; DEUTERIUM; GAS ANALYSIS; HYDROGEN; ISOTOPE RATIO; POLARIZATION; POLARIZED TARGETS
- Descriptors DEC
- ACCELERATORS; CYCLIC ACCELERATORS; ELEMENTS; HYDROGEN ISOTOPES; ISOTOPES; LIGHT NUCLEI; NONMETALS; NUCLEI; ODD-ODD NUCLEI; STABLE ISOTOPES; SYNCHROTRONS; TARGETS
Optional Information
- Copyright
- Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.