Multi-channel neutral particle energy analyser for measuring ion temperatures in tokamak plasmas
Creators
- 1. Japan Atomic Energy Research Inst., Tokai, Ibaraki. Tokai Research Establishment
Description
A multi-channel neutral particle energy analyser has been constructed for reducing the time of measurement and improving the accuracy of measurement in Tokamak plasmas. It consists of a charge stripping cell, a 10-channel electrostatic energy analyser and an ion detector. The maximum number of detected particles in the analyser is 106 cps. Energy calibration of the electrostatic energy analyser was performed with an electron beam, transmission efficiency calibration with a proton beam. The analyser was found measuring the ion temperatures in Tokamak plasmas properly. The stripping cell efficiency for H2 gas was also measured; the value is in agreement with that calculated from the charge-stripping and charge-exchange cross sections. (author)
Availability note (English)
MF available from INIS under the Report Number.Files
7220957.pdf
Files
(779.9 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:f68a6544e7cfa4a354209eedb56fd728
|
779.9 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 18 p.
- Report number
- JAERI-M--6048
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 7220957
- Subject category
- S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
- Descriptors DEI
- ACCURACY; BEAM STRIPPERS; CALIBRATION; CHARGE EXCHANGE; EFFICIENCY; ELECTRON BEAMS; ENERGY RESOLUTION; ENERGY SPECTRA; GASES; HYDROGEN; ION DETECTION; ION TEMPERATURE; MULTI-CHANNEL ANALYZERS; NEUTRAL PARTICLES; PLASMA DIAGNOSTICS; PROTON BEAMS; TEMPERATURE MEASUREMENT; TOKAMAK DEVICES; TRANSMISSION
- Descriptors DEC
- BEAMS; CHARGED PARTICLE DETECTION; CLOSED PLASMA DEVICES; ELECTRONIC EQUIPMENT; ELEMENTS; FLUIDS; LEPTON BEAMS; NONMETALS; NUCLEON BEAMS; PARTICLE BEAMS; PULSE ANALYZERS; RADIATION DETECTION; RESOLUTION; SPECTRA; THERMONUCLEAR DEVICES