Published March 1975 | Version v1
Report Open

Multi-channel neutral particle energy analyser for measuring ion temperatures in tokamak plasmas

  • 1. Japan Atomic Energy Research Inst., Tokai, Ibaraki. Tokai Research Establishment

Description

A multi-channel neutral particle energy analyser has been constructed for reducing the time of measurement and improving the accuracy of measurement in Tokamak plasmas. It consists of a charge stripping cell, a 10-channel electrostatic energy analyser and an ion detector. The maximum number of detected particles in the analyser is 106 cps. Energy calibration of the electrostatic energy analyser was performed with an electron beam, transmission efficiency calibration with a proton beam. The analyser was found measuring the ion temperatures in Tokamak plasmas properly. The stripping cell efficiency for H2 gas was also measured; the value is in agreement with that calculated from the charge-stripping and charge-exchange cross sections. (author)

Availability note (English)

MF available from INIS under the Report Number.

Files

7220957.pdf

Files (779.9 kB)

Name Size Download all
md5:f68a6544e7cfa4a354209eedb56fd728
779.9 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
18 p.
Report number
JAERI-M--6048