Published January 11, 2015 | Version v1
Journal article

First performance test of newly developed plastic scintillator for radiation detection

  • 1. Graduate School of Science and Technology, Niigata University, Niigata 950-2181 (Japan)
  • 2. Faculty of Science, Physics Department, Niigata University, Niigata 950-2181 (Japan)
  • 3. Nippon Dental University, School of Life Dentistry at Niigata, Niigata 951-8580 (Japan)
  • 4. Carlit Holdings Co., Ltd., Chiyoda, Tokyo 101-0024 (Japan)

Description

We present a plastic scintillator, developed in collaboration with Carlit Holdings Co., Ltd., that is fabricated using a liquid parent material cured at room temperature by adding a hardener. The new scintillator can incorporate heat-labile functional materials such as gadolinium to enhance neutron sensitivity. The characteristics of the new scintillator, in particular the light yield and attenuation length, were evaluated using a 90Sr β-ray source. The light yield was measured 7% Anthracene on the basis of a comparison with commercially available scintillator (BC-408) at a distance of 18 cm from the photodetector surface. This light yield is dependent on the distance between the luminous point and the photodetector because of light attenuation. The attenuation length of the Gd-doped scintillator was about 50 cm

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2014.10.005

Additional details

Identifiers

DOI
10.1016/j.nima.2014.10.005;
PII
S0168-9002(14)01122-X;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
770
Journal Page Range
p. 197-202
ISSN
0168-9002
CODEN
NIMAER