Published April 2006 | Version v1
Journal article

Distribution of the electric field intensity and seletion of parameters in the large-area flow-gas multiwire proportional counter

  • 1. China Institute for Radiation Protection, Taiyuan (China)

Description

To understand the general distribution of the electric field intensity in the large-area flow-gas multiwire proportional counter, an approximate physical model is suggested for numerical calculation and the field intensity distribution that is the relative result based on limited conditions is obtained. The relation between the distribution and some geometric parameters of the counter, such as the interval between the neighbor wires and the anode-to-cathode distance, is presented. This is advantageous for the design and application of the multiwire proportional counter. (authors)

Additional details

Publishing Information

Journal Title
Radiation Protection Bulletin
Journal Volume
26
Journal Issue
2
Journal Page Range
p. 15-20
ISSN
1004-6356

Optional Information

Notes
8 figs., 1 ref.