Published December 2015 | Version v1
Journal article

Development of a technique for improving coefficient of variation of CaSO4:Dy teflon-based TLD personnel monitoring system in low-dose region

  • 1. Radiological Physics and Advisory Division, BARC, Mumbai 400 094 (India)

Description

In view of the importance of zero-dose background (null signal) in influencing the coefficient of variation in low-dose region, a technique for the estimation of the same from composite (gross) signal is developed for CaSO4:Dy-based personnel monitoring system being used in India. The technique is based on simple analysis of glow curves (GCs) of unexposed and exposed dosemeters, evolution of trend/model for the zero-dose curves, generation of simulation protocol for individual zero-dose curves, establishment of characteristics of GCs of exposed dosemeters and finally preparation of an algorithm to segregate the components from composite signal. The technique offers the separation of real-time background and gives superior results over other method of approximation of the background. The results also prove efficiency of the empirical trending and simulation protocol of background GCs. The proposed technique can be implemented in routine monitoring without any extra man hours and reader time. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncu361

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
167
Journal Issue
4
Journal Page Range
p. 429-436
ISSN
0144-8420

Optional Information

Notes
14 refs.