Published January 1982
| Version v1
Journal article
ESR imaging spectroscopy
Description
A few coils were designed to generate magnetic field gradients for ESR Imaging and an electronic circuit was constructed for pulsation to avoid heating of coils, a cavity and its samples. The coils made of felxible print sheets were attached to the inside wall of double quartzs cooling tube inserted in a universal cavity (TE011). The coil produced a magnetic field gradient of about 100 G. cm-1 along the static magnetic field by a peak current of 10A. Another coil was constructed with copper wire to be attached to the external wall of a rectangular cavity (TE012) in order to generate a magnetic field gradient of about 50 G. cm-1 along the axis of the sample hole. (author)
Additional details
Publishing Information
- Journal Title
- Hokkaido Daigaku Kogakubu Kenkyu Hokoku
- Journal Issue
- no.107
- Series
- Hokkaido Daigaku Kogakubu Kenkyu Hokoku.
- ISSN
- 0385-602X
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 14777162
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CAVITY RESONATORS; ELECTRON SPIN RESONANCE; ELECTRONIC CIRCUITS; FABRICATION; IMAGES; MAGNET COILS; MAGNETIC FIELDS; PERFORMANCE TESTING
- Descriptors DEC
- ELECTRIC COILS; ELECTRICAL EQUIPMENT; ELECTRONIC EQUIPMENT; EQUIPMENT; MAGNETIC RESONANCE; RESONANCE; RESONATORS; TESTING