The study of the focal trough in panoramic radiograph
Creators
- 1. Department of Dental Science, Graduate School, Yonsei University, Seoul (Korea, Republic of)
Description
In the study of the focal trough of panoramic radiograph, using the Moritta company Panex EC a series of 48 exposures were taken with the 6-18 brass pins placed in the holes of the plastic model plate, then evaluated by 4 observers. The author analyzed the focal trough defined by the sharpness criteria and calculated the vertical and horizontal magnification range in the corrected focal trough. The results were as follows; 1. Continuous focal trough was not defined in the anterior region using a very high degree of sharpness. 2. As degree of sharpness used in the analysis became less, focal trough was continuous in the anterior and posterior regions, symmetrized bilaterally, and the widths of the focal trough increased more in the posterior region. 3. As sharpness criteria were reduced, the percentage range of image magnification increased in both vertical and horizontal magnification, and especially the percentage range of horizontal magnification was greater than that of vertical magnification.
Additional details
Publishing Information
- Journal Title
- Korean Journal of Oral and Maxillofacial Radiology
- Journal Volume
- 12
- Journal Issue
- 1
- Series
- 30 refs, 5 figs, 2 tabs
- Journal Page Range
- p. 115-121
- ISSN
- 1229-8212
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 42074743
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPUTER CALCULATIONS; IMAGES; MEDICAL EXAMINATIONS; PLATES
- Descriptors DEC
- MEDICAL SURVEILLANCE