Published June 2004 | Version v1
Journal article

Solid-state fluoroscopic imager for high-resolution angiography: Physical characteristics of an 8 cmx8 cm experimental prototype

  • 1. Fairchild Imaging, Inc., Milpitas, California 95035 (United States)
  • 2. Department of Radiology, Emory University School of Medicine, Atlanta, Georgia 30322 (United States)

Description

In this paper, the performance of an 8 cmx8 cm three-side buttable charge-coupled device (CCD)-based imager specially designed for high-resolution fluoroscopy and operating in fluoroscopic (30 frames/second) mode is presented in terms of the presampling modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE). The 8 cmx8 cm CCD imager is coupled to a 450 μm thick CsI:Tl scintillator by nondemagnifying (straight, 1:1) fiberoptics. The CCD imager has a fundamental pixel pitch of 39 μm and incorporates an optically opaque interline (data) channel. The CCD imager was operated at 156 μm pixel pitch by binning 4x4 adjacent pixels prior to readout. The fluoroscopic image lag was measured and accounted for in the DQE estimate to provide lag-corrected DQE. The measured limiting spatial resolution at 10% presampling MTF with the imager operated at 156 μm pixel pitch (Nyquist sampling limit: 3.21 cy/mm) was 3.6 cy/mm. In the pulsed fluoroscopic mode, the first-frame image lag was less than 0.9%. The lag-corrected DQE(0) of ∼0.62 was achieved even at a low fluoroscopic exposure rate of 1 μR/frame. Grid phantom measurements indicate no appreciable distortion. Results from DQE and image lag measurements at fluoroscopic exposure rates combined with the high spatial resolution observed from the MTF suggest that this type of imager or its variants may be a potential candidate for high-resolution neuro-interventional imaging, cardiovascular imaging, pediatric angiography, and small animal imaging. Since the CCD is three-side buttable, four such CCD modules can be joined to form a 2x2 matrix providing a field of view of 16 cmx16 cm

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
31
Journal Issue
6
Journal Page Range
p. 1462-1472
ISSN
0094-2405
CODEN
MPHYA6

Optional Information

Notes
(c) 2004 American Association of Physicists in Medicine.