Published September 2008 | Version v1
Journal article

Evaluation of Setup Error Correction for Patients Using On Board Imager in Image Guided Radiation Therapy

Creators

  • 1. Dept. of Radiation Oncology, Kosin University Gospel Hospital, Busan (Korea, Republic of)

Description

To reduce side effects in image guided radiation therapy (IGRT) and to improve the quality of life of patients, also to meet accurate SETUP condition for patients, the various SETUP correction conditions were compared and evaluated by using on board imager (OBI) during the SETUP. Each 30 cases of the head, the neck, the chest, the belly, and the pelvis in 150 cases of IGRT patients was corrected after confirmation by using OBI at every 2-3 day. Also, the difference of the SETUP through the skin-marker and the anatomic SETUP through the OBI was evaluated. General SETUP errors (Transverse, Coronal, Sagittal) through the OBI at original SETUP position were Head and Neck: 1.3 mm, Brain: 2 mm, Chest: 3 mm, Abdoman: 3.7 mm, Pelvis: 4 mm. The patients with more that 3 mm in the error range were observed in the correction devices and the patient motions by confirming in treatment room. Moreover, in the case of female patients, the result came from the position of hairs during the Head and Neck, Brain tumor. Therefore, after another SETUP in each cases of over 3 mm in the error range, the treatment was carried out. Mean error values of each parts estimated after the correction were 1 mm for the head, 1.2 mm for the neck, 2.5 mm for the chest, 2.5 mm for the belly, and 2.6 mm for the pelvis. The result showed the correction of SETUP for each treatment through OBI is extremely difficult because of the importance of SETUP in radiation treatment. However, by establishing the average standard of the patients from this research result, the better patient satisfaction and treatment results could be obtained.

Additional details

Publishing Information

Journal Title
Journal of the Korean Society for Radiotherapeutic Technology
Journal Volume
20
Journal Issue
2
Series
15 refs, 16 figs, 3 tabs
Journal Page Range
p. 69-81
ISSN
1598-8449

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
43125193
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ANATOMY; CHEST; ERRORS; EVALUATION; HEAD; IMAGES; NECK; PATIENTS; PELVIS; RADIOTHERAPY
Descriptors DEC