Published 2007 | Version v1
Journal article

Personal dosimetry in terms of HP(3):Monte Carlo and experimental studies

  • 1. INFN-LNF, Via E. Fermi 40, 00044 Frascati, Roma (Italy)
  • 2. ENEA-ION-IRP Radiation Protection Inst., Via dei Colli 16, 40136 Bologna, BO (Italy)

Description

Hp(3) has been defined as the operational quantity for eye lens dosimetry. Hp(3)/ka conversion coefficients were evaluated at the GSF (Germany) in a 30 x 30 x 15 x cm3 4-elements ICRU slab phantom for various energies and incident angles through Monte Carlo. The ISO report 12,794 suggests to employ a PMMA water filled phantom, of the same dimensions, for dosemeter calibration in terms of Hp(3). The present paper briefly summarises the main aspects of a study carried out at ENEA-Radiation Protection Inst. (Bologna (Italy)) to provide practical procedures for the calibration of dosemeters in terms of Hp(3). Tabulations of a new set conversion coefficients and air kerma backscatter factors are provided as a function of energy and incident angle. The paper demonstrates that a more accurate approach to the dosimetric assessment in terms of Hp(3) could be rather simply introduced employing a reduced phantom. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncl385

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
125
Journal Issue
1-4
Journal Page Range
p. 145-148
ISSN
0144-8420

Conference

Title
European Workshop on Individual Monitoring of Ionising Radiation
Acronym
IM2005
Dates
11-15 Apr 2005
Place
Vienna (Austria)

Optional Information

Notes
13 refs