Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.019 seconds
AbstractAbstract
[en] The dosimetry laboratory of NRCN employs a 2271 Harshaw automatic TLD card reader. The system logic performs only basic parity checks of the data which is sent to the peripheral equipment, besides the upper and lower dose limit alarms. Other errors may remain undetected until the data analysis stage. A pc computer was connected to the automatic card reader system. The pc computer controls the operation of the system while receiving, checking and recording the data on a magnetic media. A serial number is given to each record and other specifications as date, unit and type of analysis are also added to the data file. (author)
Primary Subject
Source
Alfassi, Z.B.; Ben-Gurion Univ. of the Negev, Beersheba (Israel). Dept. of Nuclear Engineering; 146 p; May 1992; p. 116-118; 17. Conference of nuclear societies of Israel; Beer Sheva (Israel); 4 May 1992
Record Type
Miscellaneous
Literature Type
Conference
Report Number
Country of publication
Reference NumberReference Number
Related RecordRelated Record
INIS VolumeINIS Volume
INIS IssueINIS Issue