[en] The assessment of simulated scintigraphic pictures by means of curves based on binary decisions of individual observers called pooled ROC-curves of pooled L-ROC-curves is presented. The technique of creating the curve is described. The examples of various processing techniques and the findings of various groups of observers are compared and discussed. The images of two IAEA simulated phantoms were used to perform the evaluation. The results indicate that the pooled L-ROC-curves may serve as a measure of image quality. (orig.) 891 MG
[de]
Es wird die Auswertung simulierter Szintigraphiebilder durch Kurven beschrieben, die auf binaeren Entscheidungen einzelner Beobachter basieren und als 'pooled ROC'-Kurven oder 'pooled L-ROC'-Kurven bezeichnet werden. Die Methode der Herstellung solcher Kurven wird beschrieben. Beispiele verschiedener Verarbeitungsmethoden und Befunde verschiedener Beobachtergruppen werden verglichen und diskutiert. Zur Auswertung wurden Bilder zweier simulierter Phantome der IAEO verwendet. Die Ergebnisse zeigen, dass die 'pooled L-ROC'-Kurven als Mass fuer die Qualitaet des Bildes dienen koennen. (orig.)Primary Subject