AbstractAbstract
[en] The purpose of this report is to define the optimum operating conditions of the whole counting systems in wavelength dispersive spectrometry (mostly used in conjunction with electron microprobes) and in energy dispersive spectrometry (more often connected to scanning electron microscopes). For both these techniques, the principle of the detector and its attached counting electronics is recalled; a check list is then given for verifying the qualities of the apparatus and detecting the defects; finally the best operating conditions are defined as essential in an analytical instrument whose reliability has to be perfect
[fr]
Ce rapport a pour but de degager les conditions optimales de fonctionnement des divers elements des ensembles de comptage en spectrometrie de rayons X a dispersion de longueur d'onde (associes generalement aux micro-analyseurs par sonde electrnique) et a selection d'energie (associes plutot aux microscopes electroniques a balayage et a transmission). Pour chacune de ces deux techniques, apres avoit fait un rappel de principe du fonctionnement du detecteur et de l'electronique qui lui est associee, on dresse un tableau des verifications a effectuer pour tester les qualites de l'appareillage et mettre eventuellement des defauts en evidence, puis on definit les meilleures conditions de reglage de ces ensembles de mesure qui sont des organes essentiels dans un instrument d'analyse dont la fiabilite doit etre parfaite durant toute sa periode d'utilisationOriginal Title
Ensembles de comptage en spectrometrie de rayons X a dispersion de longueur d'onde et a selection d'energie: principe, controle, reglage
Source
Mar 1978; 92 p
Record Type
Report
Report Number
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue