Cannon, T.M.; Fenimore, E.E.
Los Alamos Scientific Lab., NM (USA)1978
Los Alamos Scientific Lab., NM (USA)1978
AbstractAbstract
[en] Recent work in coded aperture imaging has shown that the uniformly redundant array (URA) can image distant planar radioactive sources with no artifacts. The performance of two URA apertures when used in a close-up tomographic imaging system is investigated. It is shown that a URA based on m sequences is superior to one based on quadratic residues. The m sequence array not only produces less obnoxious artifacts in tomographic imaging, but is also more resilient to some described detrimental effects of close-up imaging. It is shown that in spite of these close-up effects, tomographic depth resolution increases as the source is moved closer to the detector
Source
1978; 5 p; 22. symposium and instrument display; San Diego, CA, USA; 28 - 31 Aug 1978; CONF-780829--1; Available from NTIS., MF A01