Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.072 seconds
AbstractAbstract
[en] The organizational problems in maintaining centralized individual monitoring service with erasable and re-usable dosemeters are evaluated. Design criteria for an automated thermoluminescence reader are laid down. It is characteristic for the planning of the monitoring system that the issuing of dosemeters can be arranged without having two dosemeters for each worker. A home made reader designed to fullfil these criteria is presented. The use of a standard barcode and a standard optical barcode reader in identification of dosemeters is described. A method of using a minicomputer in preparing the self-fastening identification labels, in printing mailing lists and in printing results is described
Primary Subject
Secondary Subject
Source
p. 274-280; nd; p. 274-280; A. Scharmann; Sao Paulo; 5. International Conference on Luminescence Dosimetry; Sao Paulo, Brazil; 14 - 17 Feb 1977
Record Type
Book
Literature Type
Conference
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue