[en] In the development of fast RF amplitude measurement systems the accuracy is limited above all by the statistical errors of measurement. This limitation is smaller in a non-d.c. measurement technique as compared with a d.c. technique and is constituted by the thermal noise of the equivalent resistances. This is the first report on such a measuring system utilizing the non-linear capacity characteristic of a varactor diode to demodulate the RF amplitude. An accuracy of the amplitude measurement of 10-5 was achieved in the S band (2.7 GHz) at room temperature and a bandwith of 100 kHz. (orig.)
[de]
Die Messgenauigkeit bei der Entwicklung schneller HF-Amplitudenmesssysteme wird vor allem durch die statistischen Messfehler begrenzt. Diese Begrenzung ist bei einem gleichstromfreien Messverfahren kleiner als bei einem gleichstrombehafteten und liegt dann in dem Waermerauschen der Verlustwiderstaende. Die vorliegende Arbeit berichtet erstmals ueber ein solches Messsystem, bei dem die nichtlineare Kapazitaetskennlinie einer Varaktordiode zur Demodulation der HF-Amplitude eingesetzt wird. Bei Zimmertemperatur und einer Messbandbreite von 100 kHz konnte dabei im S-Band (2,7 GHz) eine Amplitudenmessgenauigkeit von 10-5 erreicht werden. (orig.)Original Title
Rauscharme Messung einer HF-Amplitude auf der Basis der Nichtlinearitaet einer Reaktanz-Kennlinie