Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.016 seconds
AbstractAbstract
[en] Improvements to upgrade the accuracy of Rotacon probes by a complete redesign of probe to include a Michelson interferometer to replace the existing long-range capacity transducer are described. This has resulted in a compact and interchangeable probe cartridge with a 3 μin. resolution and accuracy; the cartridge can be installed and replaced in the Rotacon gauge with the minimum of realignment, which should reduce our dependence on operator skill. In addition, the stylus contact force can be reduced to 750 mg for the contacting types, but an alternative feature, which we are still developing, will use a gas jet cushion in place of the stylus to provide a noncontacting version of the same basic probe cartridge. This device is very sensitive to external vibration effects because it is virtually frictionless
Source
Schuman, J.F. (comp.); California Univ., Livermore (USA). Lawrence Livermore Lab; p. 63-66; 1978; p. 63-66; 37. nonconvertional metrology meeting; Livermore, CA, USA; 17 - 19 Apr 1978