AbstractAbstract
[en] We reexamine the source-pulser method of measuring dead time, which is an interesting variant of the well-known two-source method. A description based on the underlying statistical processes is given which leads to an improved expression for the dead time to be determined. Thereby use has to be made of a correction factor μ discussed previously. The results permit to extend the application of the method and to check the validity of the simple approximate formula used till now