[en] A data acquisition system for experiments with the heavy ions accelerator VICKSI uses fast ADCs with high resolution. The modular CAMAC system used transfers the desired single events in a 'Single Mode' from the ADCs to the computer, while the desired coincidence events from up to 12 ADCs are transfered in a 'Group Mode'. The undesired events are rejected by hardware. This report describes the 'Single Mode' and the 'Group Mode' together with a data check and group checks, and it describes furthermore an INPUT REGISTER allowing the ADCs to work in both modes simualtanously, as well as a DUAL SHIFT INPUT REGISTER for dual channel measurements in the 'Single Mode'. (orig.)
[de]
Fuer Experimente am Schwerionenbeschleuniger VICKSI werden schnelle ADCs mit hoher Aufloesung verwendet. Das in CAMAC-Technik modular aufgebaute Messdatenerfassungssystem uebertraegt erwuenschte Einzelereignisse von ACDs im 'Single Mode' und erwuenschte Koinzidenzereignisse von maximal 12 ADCs im 'Group Mode' an den Rechner. Unerwuenschte Ereignisse weist die Hardware zurueck. Der Bericht beschreibt 'Single Mode' und 'Group Mode' mit Datenpruefung und Gruppenpruefungen sowie ein INPUT REGISTER, mit dem ADCs gleichzeitig in beiden Modes arbeiten koennen, und ein DUAL SHIFT INPUT REGISTER, welches zweiparametrische Messungen im 'Single Mode' ermoeglicht. (orig.)Original Title
ADC-Messdatenerfassungssystem fuer den Schwerionenbeschleuniger VICKSI
Primary Subject
Source
Jul 1979; 43 p