AbstractAbstract
[en] Tests on the digital individual dose ratemeter type DM 20 for gamma radiation measurement are presented. The aim of this testing is an evaluation of main physical characteristics under irradiation. Results are given in tables and graphs. Relative sensitivity versus dose rate is better than +14% for cobalt 60 photons with a good linearity. There is no detector saturation until 400 to 450 rad/h. The energy response is lower than +- 20% between 160 keV to 1.25 MeV. Autonomy is over 4000 hours. All informations: dose, dose rate and good working conditions are directly read
[fr]
On presente les essais effectues sur le dosimetre individuel numerique type DM 20, pour la mesure gamma dans le but d'evaluer les principales caracteristiques physiques du dosimetre sous irradiation. Les resultats font l'objet de tableaux et courbes. On met en evidence: une sensibilite relative en fonction du debit de dose meilleure que + 14% pour les photons du 60Co, avec une excellente linearite, une absence de saturation du detecteur jusqu'a 400 a 450 rad/heure, une reponse en energie inferieure a +- 20% entre 160 keV et 1,25 MeV, une autonomie a 400 heures. Toutes les informations de dose, debit de dose fonctionnement peuvent etre lues directementOriginal Title
Dosimetre individuel gamma DM 20: Essais physiques et essais d'evaluation
Source
Mar 1981; 12 p; Scientific and technical meeting of the French Health Physics Society; Lamanon, France; 25 Mar 1981
Record Type
Report
Literature Type
Conference
Report Number
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume