Saunders, M.T.; Drummond, C.M.; Harrison, M.A.
Australian Atomic Energy Commission Research Establishment, Lucas Heights1982
Australian Atomic Energy Commission Research Establishment, Lucas Heights1982
AbstractAbstract
[en] The Mark II molybdenum-99/technetium-99m generator now supplied to hospitals by the Australian Atomic Energy Commission is a non-sterile elution system. The Mark III version will be supplied as a sterile elution system. A validation study has been undertaken to assess the capability of the new production facility, to evaluate up-to-date procedures for manufacturing sterile generators and to demonstrate that a sterile radionuclide generator can be made. Generator manufacturing procedures and a time study of the validation are described. Microbiological methods for monitoring in-process aspects of manufacture, disinfectant efficacy and generator sterility are defined
Primary Subject
Source
Jul 1982; 20 p
Record Type
Report
Report Number
Country of publication
AUSTRALIAN ORGANIZATIONS, BETA DECAY RADIOISOTOPES, BETA-MINUS DECAY RADIOISOTOPES, DAYS LIVING RADIOISOTOPES, EVEN-ODD NUCLEI, HOURS LIVING RADIOISOTOPES, INTERMEDIATE MASS NUCLEI, ISOMERIC TRANSITION ISOTOPES, ISOTOPES, MOLYBDENUM ISOTOPES, NATIONAL ORGANIZATIONS, NUCLEI, ODD-EVEN NUCLEI, RADIOISOTOPES, TECHNETIUM ISOTOPES, YEARS LIVING RADIOISOTOPES
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue