Quam, W.; Del Duca, T.; Plake, W.; Graves, G.; DeVore, T.
EG and G, Inc., Goleta, CA (USA)1981
EG and G, Inc., Goleta, CA (USA)1981
AbstractAbstract
[en] A pocket-calculator-sized, neutron-sensitive, REM-responding personnel dosimeter is described that uses two tissue-equivelent cylindrical proportional counters as neutron-sensitive detectors. These are conventionally called Linear Energy Transfer (LET) counters. Miniaturized hybrid circuits are used for the linear pulse handling electronics, followed by a 256-channel ADC. A CMOS microprocessor is used to calculate REM exposure from the basic rads-tissue data supplied by the LET counters and also to provide timing and display functions. The instrument is used to continuously accumulate time in hours since reset, total counts accumulated, rads-tissue, and REM. At any time the user can display any one of these items or a channel number (an aid in calibration). The instrument provides such data with a precision of +-3% for a total exposure of 1 mREM over 8 hours
Source
1981; 4 p; IEEE symposium on nuclear science; San Francisco, CA (USA); 21 - 23 Oct 1981; Available from NTIS., PC A02/MF A01 as DE82017675