Davis, S.L.; Mueller, D.; Keane, C.J.
Princeton Univ., NJ (USA). Plasma Physics Lab1982
Princeton Univ., NJ (USA). Plasma Physics Lab1982
AbstractAbstract
[en] The PDX charge-exchange system is comprised of four, ten channel, mass-resolved, charge-exchange analyzers. Each analyzer is constructed with parallel electric and magnetic fields and is calibrated over an energy range of 0.5 to 40 keV. The mass rejection between hydrogen and deuterium has been measured as better than 1000 to 1. For ohmic heated discharges the system can provide single shot radial ion temperature profiles (4 point) with 1 msec time resolution. For neutral beam heated discharges complete radial and temporal profiles can be obtained in 2 to 4 shots. The system is also equipped with a vertically aimed diagnostic neutral beam to allow local ion-energy distribution measurements. This report describes the analyzer system and its calibration, and presents results from ohmic and neutral beam heated discharges
Primary Subject
Source
Oct 1982; 59 p; Available from NTIS, PC A04/MF A01 as DE83002336