AbstractAbstract
[en] Standardization of portable neutron assay instrumentation has been achieved by using the neutron coincidence technique as a common basis for a wide range of instruments and applications. The electronics originally developed for the High-Level Neutron Coincidence Counter has been adapted to both passive- and active-assay instrumentation for field verification of bulk plutonium, inventory samples, pellets, powders, nitrates, high-enriched uranium, and materials-testing-reactor, light-water-reactor, and mixed-oxide fuel assemblies. The family of detectors developed at Los Alamos National Laboratory and their performance under in-field conditions are described. 16 figures, 3 tables
Primary Subject
Secondary Subject
Source
1983; 16 p; 5. symposium on safeguards and nuclear material management; Versailles (France); 19-21 Apr 1983; CONF-830423--4; Available from NTIS, PC A02/MF A01; 1 as DE83011159