AbstractAbstract
[en] This report describes the application of the Isomass 54E Thermal ionisation mass spectrometer for the isotopic analysis of plutonium, and gives details of the development of a method for quantitative determination of plutonium in sample solutions using isotopic dilution. A computer program for the control of the 54E is also described. Isotope dilution was used to compare results of plutonium content of twelve zebra reactor fuel pellets with results obtained by titrimetry. The Isomass 54E is shown to be capable of high precision analysis of plutonium sample solutions for both isotopic content and total plutonium concentration using an isotope dilution technique. (U.K.)
Primary Subject
Source
Jun 1982; 37 p; Part 1: uranium analysis - AEEW-R--1566.
Record Type
Report
Report Number
Country of publication
ACTINIDES, BREEDER REACTORS, ELEMENTS, EPITHERMAL REACTORS, EXPERIMENTAL REACTORS, FAST REACTORS, FBR TYPE REACTORS, ISOTOPE APPLICATIONS, MEASURING INSTRUMENTS, METALS, REACTORS, RESEARCH AND TEST REACTORS, RESEARCH REACTORS, SPECTROMETERS, SPECTROSCOPY, TRACER TECHNIQUES, TRANSURANIUM ELEMENTS, ZERO POWER REACTORS