Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.014 seconds
Evaluation of response functions of a REM counter and neutron dosimeters with cylindrical moderators
Nakazawa, Masaharu; Iguchi, Tetsuo; Kohashi, Shuichi
Report of the Technical Research Committee in time of flight experiments, 1980 and 19811982
Report of the Technical Research Committee in time of flight experiments, 1980 and 19811982
AbstractAbstract
[en] The response functions of a REM counter and a neutron dosimeter were measured in the epi-thermal and the intermediate velocity regions by using a LINAC-TOF system. The efficiency curves for two types of the neutron dosimeters with moderators were measured and analyzed. The dead time correction of the time analyzer and the correction of time resolution were made for the data. The adjoint calculation by the ANISN code was made. The experimental and the calculated results were mutually compared. The preliminary works on the angle dependence of the efficiency of the detectors are in progress. As the results of this study, it was found that a method to obtain the energy dependent efficiency curves has been established. The correction of time resolution is important. The upper limit of the range of application of correction for energy is about 1 keV. (Kato, T.)
Source
Oka, Yoshiaki; Hashikura, Hiroyuki (eds.); Tokyo Univ., Tokai, Ibaraki (Japan). Nuclear Engineering Research Lab; 126 p; Feb 1982; p. 92-97
Record Type
Report
Literature Type
Numerical Data
Report Number
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue