Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.016 seconds
AbstractAbstract
[en] A commercial extrapolation chamber was tested in different β radiation fields and its properties investigated. Its usefulness for β radiation calibration and dosimetry was verified. Experiments were performed in order to obtain the main characteristics such as the calibration factors (and consequently the energy dependence) for all chamber collecting electrodes (between 10 and 40 mm diameter), the transmission factors in tissue and the useful source-detector distance range
Source
Nuclear Regulatory Commission, Washington, DC (USA). Office of Nuclear Regulatory Research; p. 137-140; Jan 1984; p. 137-140; International beta dosimetry symposium; Washington, DC (USA); 15-18 Feb 1983; Available from NTIS, PC A99/MF A01; 1 - GPO $12.00 as TI84006058
Record Type
Report
Literature Type
Conference; Numerical Data
Report Number
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue