AbstractAbstract
[en] Absolute peak detection efficiencies of a Ge(Li) detector for gamma-rays of 3.5 MeV to 12 MeV were measured using four (p,γ) reactions and a (n,γ) reaction. Two-line-method was used to obtaine peak detection efficiencies. The efficiencies with the both cases are agreed very well. Utilization of (n,γ) reaction is, therefore, effective for measuring these efficiencies, because high energy gamma-rays can be generated easily by using a neutron source. These results were applied to calibration of a gamma-ray standard source, emitting 6.13 MeV gamma-rays, and of intensities of 56Co standard gamma-ray source. (author)
Source
Nov 1985; 24 p
Record Type
Report
Report Number
Country of publication
BARYON REACTIONS, DETECTION, ELECTROMAGNETIC RADIATION, ENERGY RANGE, EVEN-EVEN NUCLEI, GAMMA RADIATION, HADRON REACTIONS, IONIZING RADIATIONS, ISOTOPES, LIGHT NUCLEI, MEASURING INSTRUMENTS, NUCLEAR REACTIONS, NUCLEI, NUCLEON REACTIONS, OXYGEN ISOTOPES, RADIATION DETECTION, RADIATION DETECTORS, RADIATIONS, RESOLUTION, SEMICONDUCTOR DETECTORS, STABLE ISOTOPES, TARGETS