AbstractAbstract
[en] The characteristics (dead-time and recovery-time) of a mono-point TSEE/OSEE Geiger counter were investigated using electronic pulse sampling techniques. After a dead-time of 12 μs, the recovery of the counting efficiency and a restoration of the initial electric field distribution occurred at time between 25 and 80 μs. (author)
Source
8. international symposium on exoelectron emission and applications; Osaka (Japan); 25-30 Aug 1985