AbstractAbstract
[en] Unrepeatability of gas quantitative mass-spectrometric analysis caused by mass variation of tracer gas added as internal standard due to changing temperature conditions in the dosing system between two experiments is estimated. The tracer dosing system consists of a large capacity V1 containing the tracer and a small capacity V2 connected to the first one via a valve. The tracer out of the V2 capacity is bleeded into the V3 analytical capacity through the second valve where there is the gas under investigation. The mixture of the latter one with tracer then immediately or after purification is bleeded into the mass-spectrometer cell. It follows from the above analysis that the most straightforward method for improving tracer dosing repeatability is to enclose the V1 and V2 capacities into a general effective thermostat
Original Title
Uluchshenie vosproizvodimosti kolichestvennogo mass-spektrometricheskogo analiza gazov
Source
For English translation see the journal Instruments and Experimental Techniques (USA).
Record Type
Journal Article
Journal
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue