[en] Precision and accuracy in the determination of alpha activity ratios of 239Pu/233U, 241Am/233U and 244Cm/233U was evaluated by preparing synthetic mixtures from solutions of enriched isotopes of 239Pu, 241Am, 244Cm and 233U. Alpha spectra were recorded by employing solid state silicon surface barrier detectors coupled to a 4K analyzer and the alpha spectra were evaluated by a method based on the geometric progression decrease for the far tail of the spectrum. Large area detector (i.e. 450 mm2) was observed to reduce the effect of nonhomogeneous distribution, if any, of the two elements present in the source. Precision and accuracy of about 1% is demonstrated. (author)
Primary Subject
Source
7 refs.; 5 tables.
Record Type
Journal Article
Journal
Journal of Radioanalytical and Nuclear Chemistry; ISSN 0236-5731;
; CODEN JRNCD; v. 106(5); p. 295-307

Country of publication
ACTINIDE NUCLEI, ALPHA DECAY RADIOISOTOPES, AMERICIUM ISOTOPES, CHARGED PARTICLE DETECTION, CURIUM ISOTOPES, DETECTION, EVEN-EVEN NUCLEI, EVEN-ODD NUCLEI, HEAVY ION DECAY RADIOISOTOPES, HEAVY NUCLEI, ISOTOPES, MATERIALS, MEASURING INSTRUMENTS, NEON 24 DECAY RADIOISOTOPES, NUCLEI, ODD-EVEN NUCLEI, PLUTONIUM ISOTOPES, RADIATION DETECTION, RADIATION DETECTORS, RADIOISOTOPES, SEMICONDUCTOR DETECTORS, SPECTROSCOPY, URANIUM ISOTOPES, YEARS LIVING RADIOISOTOPES