Derenzo, S.E.; Huesman, R.H.; Cahoon, J.L.; Geyer, A.B.; Moses, W.W.; Uber, D.C.; Vuletich, T.; Budinger, T.F.
Lawrence Berkeley Lab., CA (USA)1987
Lawrence Berkeley Lab., CA (USA)1987
AbstractAbstract
[en] We describe the imaging performance of the Donner 600-Crystal Positron Tomograph, a single 600 cm diam ring of 3 mm wide bismuth germanate (BGO) crystals coupled individually to 14 mm phototubes. With a pulse height threshold of 200 keV and a slice thickness of 5 mm, the sensitivity is 7024 eventssec per μCiml in a 20 cm cyliner of water. The measured rates for 18 μCiml are 95,000 truessec plus 20,000 randomsec. A 0.3 mm diam 22Na line source near the center of the tomograph has a circular point spread function (PSF) with a full-width at half-maximum (fwhm) of 2.6 mm. At 5 cm from the center the PSF is elliptical with a fwhm of 2.7 mm tangential /times/ 3.2 mm radial. At 10 cm the PSF has a fwhm of 2.8 mm tangential /times/ 4.8 mm radial. Attenuation data are accumulated with a 20 mCi 68Ge orbiting transmission source and 100 million coincident events are collected in 200 sec. 20 refs., 9 figs., 5 tabs
Primary Subject
Secondary Subject
Source
Oct 1987; 14 p; 34. nuclear science symposium and 19. nuclear power systems symposium; San Francisco, CA (USA); 21-23 Oct 1987; CONF-871006--48; Available from NTIS, PC A03/MF A01; 1 as DE88007375; Portions of this document are illegible in microfiche products.
Record Type
Report
Literature Type
Conference
Report Number
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume