Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.012 seconds
Kliauga, P.; Rossi, H.H.; Johnson, G.
Radiation physics, biophysics, and radiation biology: Progress report, December 1, 1987-November 30, 19881988
Radiation physics, biophysics, and radiation biology: Progress report, December 1, 1987-November 30, 19881988
AbstractAbstract
[en] This article discusses design modifications of a multi-element proportional counter. The original counter exhibited poor resolution, as measured by the width of the event-size spectrum for low-energy photons. It was also suspected that the field inside each volume was not sufficiently symmetric. Results of the modifications showed that a dramatic improvement in resolution could be obtained in the chamber with tissue-equivalent septa if their potentials were adjusted to obtain optimal resolution. The full width at half maximum then approached, although it did not equal, that of a standard spherical counter
Source
Hall, E.J.; Zaider, M.; Delegianis, M.J. (eds.); Columbia Univ., New York (USA). Radiological Research Lab; p. 63-71; Jul 1988; p. 63-71; Available from NTIS, PC A06/MF A01; 1 as DE88015151
Record Type
Report
Literature Type
Progress Report
Report Number
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue