[en] The stability check source is an integral part of a Secondary Standard Dosimetry System. The purpose of the stability check source is to confirm that the overall response of the dosimeter has not changed significantly since the instrument was calibrated. In case any change in the sensitivity of the ionization chamber or measuring assembly occurs the same is reflected in the reference check source measurements. Stability check source measurements are taken in a Primary Standard Dosimetry Laboratory (PSDL) at the time of calibration of secondary standard dosimeter and mean time (in seconds) to the reference setting of 50 scale divisions with ambient conditions of air at 20 deg. C, 101.3 kPa and 50% RH is quoted in a calibration certificate. This quoted stability check source time figure is the basis for future confirmation of overall response of the secondary standard dosimeter system. This note presents the results of stability check source measurements carried out in SSDL Pakistan over a period of five years
Primary Subject
Secondary Subject
Source
International Atomic Energy Agency, Division of Life Sciences, Vienna (Austria); World Health Organization, Geneva (Switzerland); 44 p; ISSN 1011-2669;
; Dec 1988; p. 40-42; 1 fig