[en] The bone contrast on chest radiographs obtained with computed radiography (CR, hereafter) is not decreased enough compared with that for screen/film combination, even if we use a high tube voltage. This may be because the Exposure Data Recognizer which can normalize the density and contrast of the chest images is performed. In order to increase the diagnostic accuracy for the lung disease, we propose to use a copper additional filter to decrease the bone contrast. Consequently, in order to have satisfied the bone contrast and the patient exposure dose, additional 0.3 mm copper filter and effective energy was 59.3 keV or more were required of the tube voltage of 140 kV. Further, the effective energy of 140 kV tube voltage with a 0.7 mm copper, which is the maximum thickness of the additional filter within 50 m seconds in exposure time, was 68 keV. Concluding these results, optimum range of additional copper filter is from 0.3 mm to 0.7 mm in chest high voltage radiography using CR. The bone contrast decreases sufficiently, if we use X-rays which have effective energy within this range. Our method is also useful to reduce patient dose. (author)
Primary Subject
Record Type
Journal Article
Journal
Igaku Butsuri; ISSN 1345-5354;
; v. 20(3); p. 123-130