Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.02 seconds
AbstractAbstract
[en] The neutron dose equivalent received by patients treated with high energy x-ray beams was measured in this research. A total of 13 different medical accelerators were evaluated in terms of the neutron dose equivalent in the patient plane and at the beam center. The neutron dose equivalent at the beam center was found to ranged from 0.02 to 9.4 mSv per Sv of x-ray dose and values from 0.029 to 2.58 mSv per Sv of x-ray were measured in the patient plane. It was concluded that the neutron levels meet the International Electrotechnical Commission standard for the patient plane. It was also concluded that when intensity modulated radiation treatment is conducted the neutron dose equivalent received by the patient will increase by a factor of 2 to 10. (author)
Primary Subject
Secondary Subject
Source
International Atomic Energy Agency, Vienna (Austria); European Commission, Brussels (Belgium); Pan American Health Organization, Washington, DC (United States); World Health Organization, Geneva (Switzerland); 916 p; Mar 2001; p. 605-608; International conference on radiological protection of patients in diagnostic and interventional radiology, nuclear medicine and radiotherapy; Malaga (Spain); 26-30 Mar 2001; IAEA-CN--85-8; ISSN 1563-0153;
; 6 refs, 2 tabs