Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.018 seconds
AbstractAbstract
[en] A method to establish the relationship between CT number and effective density for therapeutic radiations is proposed. We approximated body tissues to mixtures of muscle, air, fat and bone. Consequently, the relationship can be calibrated only with a CT scan of their substitutes, for which we chose water, air, ethanol and potassium phosphate solution, respectively. With simple and specific corrections for non-equivalencies of the substitutes, a calibration accuracy of 1% will be achieved. We tested the calibration method with some biological materials to verify that the proposed method would offer the accuracy, simplicity and specificity required for a standard in radiotherapy treatment planning, in particular with heavy charged particles
Primary Subject
Source
S0031-9155(03)57396-1; Available online at http://stacks.iop.org/0031-9155/48/1053/m30807.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/; Country of input: International Atomic Energy Agency (IAEA)
Record Type
Journal Article
Journal
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue