Filters
Results 1 - 1 of 1
Results 1 - 1 of 1.
Search took: 0.017 seconds
AbstractAbstract
[en] The advent of 3D treatment planning systems whose algorithms utilize diagonal scan data to perform dose calculations has made the collection of diagonal profile data essential. Manufacturers' specifications (MS) on beam flatness and symmetry apply to both the radial and transverse axes of all square field sizes from 10 x 10 cm2 to the largest field available. Beam profile measurements were obtained for both diagonal axes over a range of field sizes and depths for two units, a Varian 2100C and a Siemens KD. In this note the International Electrotechnical Commission (IEC) flatness definition was used to characterize the diagonal flatness of each beam
Primary Subject
Source
016781409601794X; Copyright (c) 1996 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: Argentina
Record Type
Journal Article
Journal
Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue