Ricq, Stephane; Glasser, Francis; Garcin, Michel, E-mail: glasser@chartreuse.cea.fr2000
AbstractAbstract
[en] The application of CdTe and CdZnTe 2D array detectors for medical X-ray Computed Tomography (XCT) is investigated. Different metallic electrodes have been deposited on High-Pressure Bridgman Method CdZnTe and on Traveling Heater Method CdTe:Cl. These detectors are exposed to X-rays in the CT irradiation conditions and are characterized experimentally in current mode. Detectors performances such as sensitivity and response speed are studied. They are correlated with charge trapping and de-trapping. The trapped carrier space charges may influence the injection from the electrodes. This enables one to get information on the nature of the predominant levels involved. The performances achieved are encouraging: dynamic ranges higher than 4 decades and current decreases of 3 decades in 4 ms after X-ray beam cut-off are obtained. Nevertheless, these detectors are still limited by high trap densities responsible for the memory effect that makes them unsuitable for XCT
Primary Subject
Source
S0168900299011985; Copyright (c) 2000 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: Pakistan
Record Type
Journal Article
Journal
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment; ISSN 0168-9002;
; CODEN NIMAER; v. 442(1-3); p. 45-52