[en] Over 10 000 channels from the CMS tile/fiber hadronic calorimeter (The CMS Hadron Calorimeter TDR, CERN/LHCC 97-31, June 20, 1997) will be read out by custom 19 and 73-channel HPD's in a 4 T field. Experimental constraints on the tube design have produced a faster tube with a larger active area and minimum accelerating gap. Quality assurance stations have been developed for fast processing of 600 tubes. Registration to 50 μm is done in situ, providing a custom collar with alignment pins which can be mated to a universal cookie. Lifetime measurements reveal details on aging, but operation is within CMS specifications
Primary Subject
Source
S016890029901236X; Copyright (c) 2000 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: Pakistan
Record Type
Journal Article
Journal
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment; ISSN 0168-9002;
; CODEN NIMAER; v. 442(1-3); p. 289-294

Country of publication
Reference NumberReference Number
INIS VolumeINIS Volume
INIS IssueINIS Issue